PH and ORP Benchtop Meter with Data Logging with Probe Each

Write a Review
SKU:
HI-2621-02
MPN:
HI-2621-02
  • PH and ORP Benchtop Meter with Data Logging with Probe Each
  • PH and ORP Benchtop Meter with Data Logging with Probe Each
  • PH and ORP Benchtop Meter with Data Logging with Probe Each
£510.00 Inc. VAT
£425.00 Ex. VAT

If you have an account:

Description

HI-2621 benchtop meter enables fast, accurate pH and ORP measurements utilising the Hanna Instruments digital electrode / probes for pH and ORP. Each digital electrode / probe has a unique serial number and the parameter that it measures is automatically identified by the meter.


Once connected, the single sensor or sensors are ready to measure the parameter along with temperature.
The intuitive design simplifies configuration, calibration, measurement, data logging and transfer of data to a USB thumb drive or computer.


The HI-2621 offers a Basic Mode that streamlines measurement configuration and is useful for routine applications as well as a Standard Operating Mode, where all features and capabilities are enabled.


The compact meter (and probe) can be used as a portable device (using its rechargeable battery) for field measurements or used as laboratory instrumentation.


The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle.

Specifications

  • Automatic parameter recognition
  • pH/ORP choice of measurement units: pH, mV;
  • CAL Check indicators
  • Sensor Check indicators
  • Basic mode for simplified operation
  • Dedicated GLP key
  • GLP data included with logged data
  • Simplified data transfer to a PC
  • Internal clock and date

Key Features

  • Digital probe input - simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.
  • Clear, full text readout - The HI-2621 presents clear, detailed text prompts at the bottom section of the screen.
  • Two operating modes - HI-2621 can be used in Standard or Basic Operating Modes (pH and EC only). Standard Mode enables all features while Basic Mode displays a simplified screen and features —ideal for routine measurements.
  • Data logging - The HI-2621 provides ample data storage capacity, allowing you to save up to 1000 log records. Each data set comprehensively includes readings, GLP information, as well as date and time stamps.
  • GLP - The sensor retains data from the most recent calibration, including date, time, and buffers utilised. When connected to the HI-2621, this GLP information is automatically transferred.
  • CAL Check - Hanna’s exclusive CAL Check feature analyses the pH electrode’s response in calibration buffers to detect potential issues such as contaminated solutions or dirty electrodes. Upon completing calibration, indicators reflecting the probe’s condition—based on its offset and slope characteristics—are displayed on the measurement screen.
  • Sensor Check - When paired with compatible Hanna electrodes featuring a matching pin, HI-2600 continuously monitors the impedance of the pH measuring electrode, providing real-time alerts in case of glass breakage. During calibration, Sensor Check displays the junction condition, by evaluating the reference junction resistance.
  • ORP measurement - Measure ORP (REDOX) in mV units by using a compatible ORP sensor.
  • Capacitive touch keypad - HI-2621 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue.
  • Easy to read LCD - HI-2621 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’). The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry.
  • USB-C port for charging and data transfer - The HI-2621 features a streamlined USB-C port that facilitates transferring data to your computer.

Key Features

  • HI-11310 pH digital electrode
  • HI-764026 Electrode holder with the following accessories: base plate with integrated pivot pin; cable holder clip, attached
  • pH calibration kit: pH 4 buffer solution (2 sachets); pH 7 buffer solution (4 sachets); pH 10 buffer solution sachets (2 sachets)
  • Electrode cleaning solution (2 sachets)
  • HI-920018 - 1.5m USB-C to USB-C cable
  • USB-C power adapter
  • Instrument quality certificate
  • Probe / electrode quality certificate
  • Battery safety insert
  • Quick reference guide with QR code for instruction manual download

Download Brochure

View AllClose

Additional Information

Temperature Range:
-20.0 to 120.0 °C (temp limits reduced to probe limits)
Dimensions:
205 x 160 x 77 mm
Weight:
Approximately 0.85 kg
View AllClose